Medical Device Certifications & Regulatory Requirements

What the main standards, conformity marks, approvals, registrations and licences mean, and which apply in 20 markets. Search an entry or browse by category below.

Marks, standards and authorities at a glance

Badges are text labels made by SurgiFacts for navigation. They are not official logos or marks and do not imply endorsement, certification or approval by any body.

Quality management & ISO standards

Standards for how a company runs its quality system and manages risk. A QMS certificate covers the organisation, not a product, and is not a market authorization.

Product safety & technical standards

Technical standards describe how to design, test or validate a device. Meeting one is evidence toward legal requirements; it is not itself an approval or a certificate to sell.

International regulatory authorities

Regulators grant the legal permission to supply a device. Depending on the market and risk class that is a clearance or approval, a registration or listing, and often a licence for the company itself.

Regional conformity frameworks

Conformity assessment shows a device meets the rules. A conformity marking is affixed by the manufacturer after that assessment; it is not a certificate issued to the product by a regulator.

Country and regional requirements

The framework and main requirements for each market, after the international standards. Open an entry for scope, typical documentation, process and official sources. Related: Surgical instrument manufacturers · Medical AI companies

Filters

Clear all
Country or region
Requirement type
Device category
Applicability

7 entries

Pakistan

PakistanImporter or Distributor RequirementsLabeling and TraceabilityManufacturer or Establishment Licensing

Pakistan Medical Device Regulatory Framework (DRAP Act 2012 and Medical Devices Rules, 2017) (DRAP MDR 2017)

Drug Regulatory Authority of Pakistan (DRAP) – Medical Device Board (MDB), Division of Medical Devices and Medicated Cosmetics

Pakistan regulates medical devices under the Medical Devices Rules, 2017: establishments need a DRAP licence and each device needs enlistment (Class A) or registration (B–D).

Purpose

The Medical Devices Rules, 2017 (notified 16 January 2018 under the DRAP Act, 2012) set up a two-layer system: licensing of the establishment that manufactures or imports, and enlistment or registration of each device. The Medical Device Board (MDB) of DRAP is the decision-making body.

Who it applies to

Local manufacturers and importers placing medical devices, components or accessories on the Pakistani market, and establishments exporting devices (export permits).

Applicable products

Instruments, apparatus, implants, in-vitro reagents, software and similar articles with a medical purpose, excluding products whose primary action is pharmacological, immunological or metabolic.

Classification

Four risk classes: A (low), B (low–moderate), C (moderate–high), D (high), assigned using the rules in Schedule A of MDR 2017 (GHTF/IMDRF-based).

General requirements

  • Establishment licence to manufacture or to import (separate licence per premises)
  • Enlistment of Class A devices; registration of Class B, C and D devices
  • Applications for enlistment/registration can only be made by a licensed local manufacturer or a licensed importer acting as authorized representative
  • Labelling, warranty (Form-5) and record-keeping conditions attached to enlistment/registration
  • Adverse event and safety-action reporting to the MDB
  • Export permit for exporting medical devices

Typical documentation

  • Application forms and checklists published by DRAP for each licence/enlistment/registration type
  • Proof of fee deposit as per Schedule C
  • Manufacturer credentials, QMS certificates and technical information (see separate entries)

A general list; what is actually required depends on the device, its class and the manufacturer’s role.

Assessment process

Obtain the establishment licence first (Form-1 manufacture / Form-2 import), then apply for enlistment (Form-6/6A) or registration (Form-7/7A) of each device through DRAP's online system with hard-copy submission where required.

Validity and renewal

Establishment licences and enlistment/registration certificates are valid for five years unless suspended or cancelled (MDR 2017, rules 5 and 15).

Important notes

  • Establishment licensing (who you are/where you operate) and product enlistment/registration (what you sell) are separate approvals; both are needed to import or manufacture for sale.
  • The Rules have been amended several times (e.g. S.R.O. 526(I)/2021, 1500(I)/2021, 430(I)/2022, 559(I)/2022, 224(I)/2023, 1049(I)/2025); always check DRAP's consolidated version.
  • DRAP's GDP guidance for medical devices and the storage-inspection checklist apply to importers' premises.

Points still being verified

  • Rule text reviewed from the 2018 notified PDF; several later SRO amendments exist (incl. S.R.O. 1049(I)/2025 amending rule 19) whose full text was not reviewed – confirm against DRAP's consolidated rules.
  • DRAP lists a separate S.R.O. 1450(I)/2026 Medical Devices (Refurbishment) Regulations, 2026 – not reviewed.
PakistanImporter or Distributor RequirementsManufacturer or Establishment LicensingQuality Management Standards

Establishment Licence to Manufacture or Import Medical Devices (MDR 2017)

DRAP – Medical Device Board (MDB)

DRAP licence needed by every Pakistani manufacturer or importer of medical devices before any device can be enlisted, registered, imported or made for sale.

Purpose

Confirms that the premises, equipment, staff and storage of a manufacturer or importer are suitable before it handles medical devices.

Who it applies to

Local manufacturers of medical devices, components or accessories, and importers (who also act as the authorized representative of foreign manufacturers).

Applicable products

All classes of medical devices, components and accessories. A separate licence is issued for each manufacturing or import premises; declared distribution godowns in federal/provincial capitals can be covered under the licence.

Classification

Not class-dependent; two licence types: manufacture and import.

General requirements

  • Suitable premises, space, plant and equipment for the intended activity (rule 6 conditions)
  • Named qualified technical person(s) to supervise manufacture or import/storage
  • Storage meeting DRAP good storage / good distribution practice expectations for medical devices
  • Possible inspection of premises by an MDB panel of experts before grant
  • Compliance with licence conditions throughout validity

Typical documentation

  • Application on Form-1 (manufacture) or Form-2 (import), via DRAP's online licensing system plus hard copy
  • Fee as per rule 63 / Schedule C
  • Details of premises, qualified technical persons and partners/directors, with affidavits as required by the form
  • DRAP checklist for grant or renewal of the relevant licence

A general list; what is actually required depends on the device, its class and the manufacturer’s role.

Assessment process

Apply online and submit the hard-copy form; DRAP may inspect the premises (importers against the Good Storage Practices checklist); the MDB then grants the licence (Form-3 manufacture / Form-4 import).

Validity and renewal

Five years from issue unless suspended or cancelled (rule 5(3)); renewal under the rules.

Important notes

  • The licence does not authorize sale of any specific device – each device still needs enlistment or registration.
  • Foreign manufacturers cannot apply directly; their Pakistani importer holds the licence and applies for product enlistment/registration.

Points still being verified

  • DRAP's GDP guideline for medical devices dates from 2016; check whether a newer version applies.
  • MDR 2017 does not appear (in the 2018 text) to create a separate wholesale/retail licence for devices outside Schedule D; confirm whether provincial or later rules add distributor licensing.
PakistanProduct Registration or ListingRegulatory Approval

Medical Device Enlistment (Class A) and Registration (Class B, C, D)

DRAP – Medical Device Board (MDB)

Each device marketed in Pakistan must be enlisted (Class A, Form-6/6A) or registered (Class B–D, Form-7/7A) by a DRAP-licensed manufacturer or importer.

Purpose

Product-level authorization confirming that a specific device meets DRAP's safety, performance and quality requirements before sale.

Who it applies to

Licensed local manufacturers (Form-6 / Form-7) and licensed importers acting as authorized representative of the foreign manufacturer (Form-6A / Form-7A).

Applicable products

All medical devices. Class A = enlistment; Classes B, C and D = registration. Devices may be grouped per Schedule B rules.

Classification

Enlistment for Class A; registration for Class B, C and D (rule 13).

General requirements

  • Valid establishment licence of the applicant
  • Application on the correct form with fee and documents/samples
  • Manufacturing methods and testing procedures must follow principles of validation
  • Labelling and prescribing information as approved in the country of origin or in the EU, USA, Japan, Canada or Australia
  • After approval: approved licence/registration number and MRP shown on labelling; changes need MDB approval

Typical documentation

  • Form-6/6A or Form-7/7A and the DRAP checklist
  • Notarized letter of authorization/agency agreement from the manufacturer (imports)
  • Free sale certificate from country of origin and/or reference country, attested as required
  • ISO 13485 / GMP certificate (and full QA or design-examination certificate where applicable), notarized
  • Declaration of conformity and essential principles checklist
  • Labels, IFU and further technical data for devices not approved in reference countries

A general list; what is actually required depends on the device, its class and the manufacturer’s role.

Assessment process

Submit the application to the MDB; DRAP may inspect the establishment (inspection may be waived where reliance applies); if satisfied, the MDB enters the device in the register and issues a certificate (Form-8/8A) with an enlistment/registration number.

Validity and renewal

Five years from enlistment/registration (rule 15(4)); renewal is applied for under rule 16, with a late-renewal window of up to one year with additional fee.

Important notes

  • Rule 15(2) allows registration to rely on approvals from listed reference regulators (USA, Japan, Australia, Canada, several European countries, UK), WHO prequalification or CE marking by a NANDO-listed notified body.
  • S.R.O. 224(I)/2023 gave licensed importers/manufacturers temporary exemption from enlistment/registration (Classes C and D until 31 Dec 2023; Classes A and B until 31 Dec 2024). These dates have passed – assume enlistment/registration is now required unless DRAP has issued a further extension.
  • Certificates cannot be transferred to another person without MDB approval.

Points still being verified

  • Status after the S.R.O. 224(I)/2023 exemption deadlines (31 Dec 2023 / 31 Dec 2024) not confirmed on an official DRAP page; check for any later extension notifications.
  • Fees exist in Schedule C but were deliberately not reproduced; confirm current amounts on DRAP's Regulatory Fees page.
PakistanRisk Management Standards

Pakistan Medical Device Risk Classification (Schedule A, MDR 2017)

DRAP – Medical Device Board (MDB)

Pakistan classes devices A–D by risk using Schedule A rules; reusable surgical instruments not connected to an active device fall in Class A.

Purpose

Sets the risk class that decides whether a device is enlisted or registered and how much evidence is needed.

Who it applies to

Manufacturers and importers, who must state the class and the classification rule applied in their application.

Applicable products

All medical devices including IVDs (IVD classification and device grouping are in the Schedules).

Classification

Class A low hazard (e.g. tongue depressors), Class B low–moderate (e.g. hypodermic needles, suction equipment), Class C moderate–high (e.g. lung ventilators, bone fixation plates), Class D high (e.g. heart valves, implantable defibrillators).

General requirements

  • Apply Schedule A classification rules (GHTF/IMDRF-based) considering intended use, duration of contact, invasiveness and whether the device is active
  • Use Schedule B rules when grouping devices in one application
  • Declare the class with the relevant rule in the enlistment/registration application

Typical documentation

  • Classification justification citing the Schedule A rule

A general list; what is actually required depends on the device, its class and the manufacturer’s role.

Assessment process

The manufacturer/importer determines the class; disputes can be referred to the MDB for decision.

Important notes

  • MDR 2017 defines a 'reusable surgical instrument' as one used for cutting, drilling, sawing, scratching, scraping, clamping, retracting, clipping etc., not connected to an active device, and intended to be reused after cleaning, disinfection or sterilization.
  • Schedule A Rule 6: surgically invasive devices for transient use are Class B unless they are reusable surgical instruments (Class A); a surgical instrument connected to an active device is above Class A. Rule 6 has further exceptions (e.g. energy-supplying devices) that move devices higher.
  • Single-use surgical instruments (e.g. single-use scalpels) are given as Class B examples under Rule 6.
PakistanBiocompatibilityLabeling and TraceabilityProduct Registration or Listing

Technical Documentation for Medical Device Registration in Pakistan

DRAP – Medical Device Board (MDB)

DRAP's Form-7A checklist lists the dossier for imported Class B–D devices: manufacturer credentials, ISO 13485/GMP, free sale, DoC, labels, plus technical data if not RRA-approved.

Purpose

Explains what evidence DRAP generally expects in an enlistment/registration dossier.

Who it applies to

Importers (Form-6A/7A) and local manufacturers (Form-6/7) preparing applications.

Applicable products

Content varies by class and form; the deepest technical section applies to devices not approved or on free sale in the reference countries listed in the Rules.

Classification

Applies to all classes; checklist content differs between Class A enlistment and Class B–D registration.

General requirements

  • Manufacturer details including all sites and sub-assembly processes
  • Product details: brand/generic name, HS/GMDN code, class and rule, manufacturing/QC summary, shelf life and storage with stability justification, proposed MRP
  • Statement whether the device is for export or local market
  • Grouping information, list of configurations, accessories and intended use

Typical documentation

  • Notarized agency agreement/letter of authorization naming the importer
  • Free sale certificate from the country of origin attested by the Embassy of Pakistan, and/or free sale certificate from a reference regulator
  • ISO 13485 / GMP certificate; full QA or design-examination certificate where applicable (notarized)
  • Essential principles of safety and performance; declaration of conformity on manufacturer letterhead
  • Label, packaging and IFU as approved in the country of origin
  • For devices not approved in reference countries: contraindications/warnings, sterilization/measuring validation, software validation (active devices), biological-material and biocompatibility data, pre-clinical test reports
  • Declaration on stamp paper and a soft copy of the application

A general list; what is actually required depends on the device, its class and the manufacturer’s role.

Assessment process

Compile the dossier against the DRAP checklist for the relevant form and submit with the application.

Important notes

  • This list is from the official DRAP checklist for imported Class B–D devices; Class A and local-manufacture checklists are separate and shorter.
  • Notarization and embassy attestation requirements apply to foreign documents.

Points still being verified

  • Checklist version is from 2022 uploads; Form-6A/7A were amended by S.R.O. 430(I)/2022 – confirm the current checklist on DRAP's application-forms page.
PakistanLabeling and TraceabilityPost-Market Surveillance

Post-Market Obligations for Enlisted/Registered Medical Devices in Pakistan

DRAP – Medical Device Board (MDB)

Pakistani enlistment/registration holders must keep safety records, report failures and side-effects to DRAP within set deadlines, and report foreign safety actions.

Purpose

Keeps DRAP informed of safety and quality problems after a device is on the market so it can act (e.g. recalls).

Who it applies to

Enlistment/registration holders: local manufacturers and importers.

Applicable products

All enlisted or registered devices; record-keeping duties in rule 19(8) are framed for 'new' medical devices.

General requirements

  • Keep organized records of clinical data, literature, labelling errors, batch deterioration, batch failures, unexpected side effects and performance failures (rule 19(8))
  • Report batch failures, unexpected side effects/injuries and unusual performance failures to the MDB as soon as possible and within 14 days (rule 19(9)(b))
  • Inform the MDB within 30 days if the device is withdrawn, banned or restricted for safety reasons in the origin country or the USA, EU, Canada, Japan or Australia (rule 19(10))
  • Report within 30 days any adverse reaction reported in reference countries that was not in the application (rule 19(12))
  • Importers must ensure regular, adequate supply; issue warranty on Form-5

Typical documentation

  • Complaint and adverse-event records; recall procedure under DRAP GDP guidance

A general list; what is actually required depends on the device, its class and the manufacturer’s role.

Assessment process

Reports go to the MDB; DRAP also runs a National Pharmacovigilance Centre that accepts adverse-event reports and issues recall alerts for substandard devices.

Important notes

  • MDR 2017 defines 'field corrective action' as action by the manufacturer or its authorized agent to reduce a risk of death or serious deterioration in health.

Points still being verified

  • S.R.O. 1049(I)/2025 amended rule 19; its text was not reviewed, so these deadlines may have changed.
PakistanManufacturer or Establishment LicensingProduct Registration or Listing

Export Permit for Medical Devices (MDR 2017, rules 27–36)

DRAP – Medical Device Board (MDB)

Under MDR 2017 a Pakistani establishment needs MDB approval and an export permit (Form-12/13) to export devices, and must be licensed with devices enlisted/registered.

Purpose

Controls export of medical devices from Pakistan, which is relevant to export-oriented instrument makers.

Who it applies to

Establishments exporting medical devices from Pakistan (except small quantities for personal use or for clinical investigation/testing, which have their own permits).

Applicable products

Medical devices covered by the Rules, including reusable surgical instruments (Class A).

Classification

Not class-dependent.

General requirements

  • No export without MDB (or authorized officer) approval (rule 27(1))
  • No export permit unless the establishment is licensed and the device is enlisted or registered (rule 27(2))
  • Keep records of all exports; allow inspection of storage and sampling; withdraw batches from export if directed (rules 31–32)
  • Consignments to be accompanied by invoice/other documents as specified (rule 35)

Typical documentation

  • Application for export permit on Form-12 with fee per rule 63

A general list; what is actually required depends on the device, its class and the manufacturer’s role.

Assessment process

Apply on Form-12; one application and one permit can cover several devices from the same manufacturer; the permit is issued on Form-13.

Validity and renewal

Export permit valid for three years unless suspended or cancelled; it stays valid pending renewal if renewal is applied for three months before expiry (rule 30).

Important notes

  • The forms ask whether a device is 'for export only', but in the 2018 text reviewed, export still required the establishment to be licensed and the device enlisted/registered. No separate export-only exemption was found in DRAP sources reviewed.
  • Exporters also need to meet the importing country's own rules (e.g. EU MDR, US FDA).

Points still being verified

  • Export provisions read from the 2018 notified text only; check later amendments and current DRAP practice for export-only manufacturers (e.g. Sialkot instrument makers).
Official sources (1)

Last verified 29 October 2026

Understanding certifications

These terms are not interchangeable. A quality-system certificate, a product authorization, a registration and an establishment licence are different things, granted by different bodies for different purposes. Which ones apply depends on the product, its intended use, its risk class, the company’s activities and the target market.

Product certification

Product certification is a third-party attestation that a specific product meets stated requirements; it is not the same as permission to sell a medical device.

Explains what a product certificate or certification mark means: an independent body has evaluated a product (or product type) against defined requirements, such as a safety standard, and attests that it conforms.

  • A product certificate does not by itself authorize sale of a medical device in a regulated market; regulators have their own authorization routes.
  • Product certification is different from QMS certification (e.g. ISO 13485), which covers the organization's management system, not a specific product.
  • Check that a certificate names the exact model, the standard edition, and the issuing body; certificates cannot be transferred to look-alike products.
  • Example: CE marking of a medical device is a manufacturer's declaration of conformity with EU law, sometimes involving a notified body; it is not a voluntary private certification mark.

Regulatory approval / marketing authorization

Marketing authorization is a regulator's (or legally designated body's) decision allowing a device to be placed on a market; routes and names differ by country.

Explains the legal step that permits a medical device to be sold in a given jurisdiction, and why different terms (clearance, approval, certificate, licence, registration) are not interchangeable.

  • FDA 510(k) clearance, De Novo grant and PMA approval are different outcomes; calling a cleared device 'FDA approved' is inaccurate.
  • FDA establishment registration and device listing are not FDA approval of the establishment or its devices.
  • An establishment licence (permission to operate a business activity) is different from authorization to market a specific device.
  • Authorization in one country does not carry over to another, although some regulators use reliance or abridged routes.

Product registration and listing

Registration and listing put a company or device on a regulator's record; in many systems this is administrative and does not mean the device was reviewed or approved.

Explains that many regulators keep databases of establishments and devices, and that the meaning of 'registration' varies widely: in some countries it is a filing, in others it is the actual marketing authorization.

  • US: FDA establishment registration and device listing are required for many firms, but FDA states that registration does not denote approval of the establishment or its devices.
  • EU: actor and device registration in EUDAMED is a separate obligation from conformity assessment and CE marking.
  • In some countries 'registration' is the name of the marketing authorization itself, so always read what the specific regulator means.
  • A registration number alone is not evidence of quality, safety or performance.

Manufacturer / establishment licensing

An establishment licence permits a company to carry out activities such as manufacturing, importing or distributing devices; it does not authorize any specific device.

Separates permission for a business to operate (a site or activity licence) from permission to market a particular device.

  • Holding an establishment or manufacturing licence does not mean every product the company makes is authorized for sale.
  • Conversely, a device authorization usually does not license an unrelated importer or distributor to handle it.
  • Names differ: licence, permit, registration, manufacturing licence, import licence.

Quality management system (QMS) certification

QMS certification confirms an organization's management system meets a standard such as ISO 13485; it does not certify or authorize any individual product.

Explains what an ISO 13485 or ISO 9001 certificate means: an independent certification body audited the organization's QMS against the standard for a stated scope and sites.

  • ISO itself does not certify organizations; certification bodies do.
  • ISO 13485 is a QMS standard; an ISO 13485 certificate is not a product approval, CE marking or FDA clearance.
  • Some regulators require or rely on QMS certification (e.g. Health Canada requires an MDSAP-issued ISO 13485 certificate for Class II-IV licence applications), which is a regulatory use of a voluntary standard.
  • Check that the certification body is accredited for that scope and that the certificate can be verified with the issuer.
  • QMS certification is different from FDA inspection: FDA does not issue ISO 13485 certificates.

Conformity assessment

Conformity assessment is any process showing that specified requirements are met: testing, inspection, certification, or a manufacturer's own declaration.

Gives the umbrella term for how products, processes, systems or organizations are shown to meet requirements, and explains its legal meaning in the EU medical device framework.

  • CE marking signals that the manufacturer claims conformity with all applicable EU requirements following the required conformity assessment; it is not a quality mark or a product certification by the EU.
  • A notified body certificate under the MDR is part of conformity assessment, not an EU 'approval' issued by the European Commission.
  • Using a harmonised standard gives a presumption of conformity only for the requirements that standard covers.

Accreditation of certification, inspection and testing bodies

Accreditation is a formal check of a certification body or laboratory's competence; it is not a certification of the device maker.

Explains the layer above certification: accreditation bodies assess whether certification bodies, labs and inspection bodies are competent and impartial.

  • Manufacturers are certified, not accredited; a company claiming to be 'ISO 13485 accredited' is usually misusing the term.
  • Regulatory designation is separate from accreditation: EU notified bodies are designated by member-state authorities under the MDR/IVDR, and MDSAP auditing organizations are recognized by the participating regulators.
  • FDA's ASCA programme uses accredited testing laboratories for certain recognized standards; it is a voluntary FDA programme in which testing laboratories accredited to ISO/IEC 17025 plus ASCA specifications perform testing for premarket submissions; it does not replace premarket review.
  • Unaccredited certificates may not be accepted by buyers or regulators.

Product safety and performance standards

Safety and performance standards define technical methods and requirements for products; they are voluntary unless a law or regulator makes them mandatory or recognises them.

Explains how technical standards (e.g. IEC 60601 series, ISO 10993 series, ISO 11135/11137) relate to regulation: they are tools to show requirements are met, not laws themselves.

  • In the EU, applying harmonised standards is voluntary but gives a presumption of conformity with the requirements they cover.
  • FDA recognition can be complete or partial; a declaration of conformity to a recognized standard can support, but does not replace, a premarket submission.
  • Standards have editions and transition periods; an outdated edition may no longer be accepted.
  • Not every standard applies to every device: e.g. IEC 60601-1 applies to medical electrical equipment, not to a manual stainless-steel forceps.

Market-specific labelling and documentation

Each market sets its own rules for labels, instructions for use, languages, identifiers and local contact details; international standards help but do not override them.

Explains why the same device may need different labels and documents in different markets, and how labelling standards fit in.

  • ISO 15223-1 symbols and ISO 20417 content requirements are voluntary standards; the law of each market decides what is mandatory.
  • EU MDR Annex I Chapter III sets EU labelling and IFU requirements; FDA labelling rules are in 21 CFR (e.g. Part 801).
  • A symbol accepted in one market may need accompanying text in another.

Medical device classification

Every market classifies devices by risk, but the systems differ: US Class I/II/III, EU Class I/IIa/IIb/III and India’s Class A–D are not equivalent. Classification follows the intended purpose and each jurisdiction’s rules, not the manufacturer’s product description alone.

Risk-based device classification

Devices are grouped into risk classes that drive the authorization route; each country has its own rules and the class systems are not interchangeable.

Explains that regulators sort devices from lower to higher risk, and that the class determines how much evidence and third-party involvement is needed.

  • US uses Class I, II, III; the EU MDR uses Class I, IIa, IIb, III (with sub-categories such as sterile, measuring and reusable surgical instruments in Class I); India's Medical Devices Rules 2017 use Classes A, B, C, D. These do not map one-to-one.
  • Classification depends on intended purpose, duration of contact, invasiveness, body site, energy source and other factors defined in each jurisdiction's rules.
  • A manufacturer's marketing description or a supplier's catalogue label does not decide the class; the applicable rules applied to the intended purpose do.
  • The same instrument can fall into different classes in different markets, and a change in intended use can change the class.
  • IVD classification is separate in many systems (e.g. EU IVDR Classes A-D).

Related FAQs

All certification and regulatory FAQs

This directory is provided for general informational and educational purposes only. Medical device regulatory requirements vary according to product classification, intended use, manufacturer or importer activities, and destination market. Requirements may change over time. This information is not legal or regulatory advice and does not guarantee market authorization. Manufacturers should verify the applicable requirements with the relevant regulatory authority and qualified regulatory professionals.